• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSIGHTEC EXABLATE 4000; MR GUIDED FOCUSED ULTRASOUND SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSIGHTEC EXABLATE 4000; MR GUIDED FOCUSED ULTRASOUND SYSTEM Back to Search Results
Model Number 4000
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Paralysis (1997); No Code Available (3191)
Event Date 05/21/2019
Event Type  Injury  
Manufacturer Narrative
The retrospective analysis has not indicated any technical failures of the system.Treatment parameters were in line with typical range.The system performance was found to be to spec and as expected.
 
Event Description
Patient underwent brain treatment of tremor-dominant parkinson's disease.7 days after treatment the patient reported motor left sided hemiparesis, spasticity left leg, painful dystonia distal left upper limb, tremor recurrence.On july 6th, around two weeks post treatment, the resident neurologist reported a mild improvement on some of the adverse effect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXABLATE 4000
Type of Device
MR GUIDED FOCUSED ULTRASOUND SYSTEM
Manufacturer (Section D)
INSIGHTEC
nachum heth 5
tirat carmel, israel 39120
IS  39120
Manufacturer (Section G)
INSIGHTEC
nachum heth 5
tirat carmel, 39120
IS   39120
Manufacturer Contact
meital trank
5 nachum heth
tirat carmel,, 39120
IS   39120
MDR Report Key8815647
MDR Text Key151850047
Report Number9615058-2019-00009
Device Sequence Number1
Product Code POH
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P150038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number4000
Device Catalogue NumberSYS940351
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/26/2019
Initial Date FDA Received07/23/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Disability;
-
-