A patient information was not provided.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).However investigation of similar events showed the issue is more likely associated to a hole inside the evaporator, that allows the fluid to enter the cooling circuit.In case of a hole of the evaporator, the refrigerant fluid will leak and water will enter the cooling circuit.This situation will cause a damage of the cooling compressor.The compressor failure leads to an increase of the leakage current of the device.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Corrective actions are in progress for this issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|