The initial reporter complained of a software issue on a cobas u 601 urine analyzer with software version 2.2.7 which could impact patient results, interpretation of patient results, or patient identification.The customer stated that the same patient identification for the same sample rack and position was used for two different patient results ran at different times.The customer stated that the same sample was not rerun.At 16:11 the results from the u 601 were color was pale yellow, clarity was clear, ph was 7, specific gravity was 1.031, nitrate was negative, protein was negative, glucose was 4+, ketone was negative, urobilinogen was negative, bilirubin was negative, erythrocytes were negative, and leukocytes were 3+.At 16:38 the results from the cobas u 601 for the same sample identification from the same sample rack and position were color was pale yellow, clarity was clear, ph was 6.5, specific gravity was 1.012, nitrate was negative, protein was negative, glucose was negative, ketone was negative, urobilinogen was negative, bilirubin was negative, erythrocytes were negative, and leukocytes were negative.No questionable results were reported outside of the laboratory.There was no allegation of an adverse event.No reagent lot numbers or expiration dates were provided.
|