We have received the complaint device for evaluation and we have confirmed the reported incident.The balloon had dislodged from its catheter shaft.The dislodged balloon had its distal and proximal ligature intact on it.We did not observe any necking of the extrusion.We have conducted a lot history review for this lot number and did not find any issues noted in the manufacturing or packaging process that could be related to this issue.All qc tests passed their requirements.Further, we have not received any other complaints of a similar nature for devices from this lot.At this time, we could not conclusively determine the root cause of the defect.It is possible that the tension or the amount of glue applied underneath the balloon during winding of the threads to the lumen of the catheter was not strong enough to prevent the ligature from slipping from its position.Device was tested by the surgeon before use and was found to be functional.During the procedure, surgeon had attempted to remove the thrombus multiple times before encountering the issue.It is also possible that the tortuous anatomy of the patient, calcified plaque that were present in the patient's vessel, and the tension on the device during the procedure could have also led to this failure.Please note that our manufacturing team does perform 100% inspection on each device to ensure that the ligature is properly bound to the shaft of the catheter.The arterial embolectomy catheter is not recommended for the removal of fibrous, adherent or calcified material ( eg.Chronic clot, atherosclerotic plaque ) since this catheter is not designed to withstand the additional pull force needed to remove these materials.Surgeon was able to remove the dislodged balloon from the patient's vessel.There was no injury to the patient as a result of this incident.
|