• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Catalog Number SGC0301
Device Problems Leak/Splash (1354); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/03/2019
Event Type  malfunction  
Manufacturer Narrative
The customer reported the device was discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.(b)(4).
 
Event Description
This is filed to report a leak.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.Prior to the procedure it was reported that an inferior vena cava (ivc) filter had been previously implanted.While advancing the steerable guide catheter (sgc), difficulty was experienced crossing the ivc filter.The sgc was removed and a balloon was used in an attempt to create room for the sgc to pass through.The sgc was re-inserted but was still unable to cross the ivc filter.The sgc was removed and attempted to be re-flushed; however, a loss of fluid column was observed.The sgc was replaced and two clips were implanted, reducing mr to a grade of 1.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar incidents reported for leaks and difficult to advance from this lot.All available information was investigated and the reported difficult to advance the steerable guide catheter (sgc) appears to be due to procedural circumstances; however, a definitive cause for the reported leak (loss of fluid column) could not be determined in this incident.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key8822276
MDR Text Key152069223
Report Number2024168-2019-10296
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648216824
UDI-Public08717648216824
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/25/2020
Device Catalogue NumberSGC0301
Device Lot Number90326U154
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/03/2019
Initial Date FDA Received07/24/2019
Supplement Dates Manufacturer Received08/15/2019
Supplement Dates FDA Received08/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-