This is filed to report a leak.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.Prior to the procedure it was reported that an inferior vena cava (ivc) filter had been previously implanted.While advancing the steerable guide catheter (sgc), difficulty was experienced crossing the ivc filter.The sgc was removed and a balloon was used in an attempt to create room for the sgc to pass through.The sgc was re-inserted but was still unable to cross the ivc filter.The sgc was removed and attempted to be re-flushed; however, a loss of fluid column was observed.The sgc was replaced and two clips were implanted, reducing mr to a grade of 1.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar incidents reported for leaks and difficult to advance from this lot.All available information was investigated and the reported difficult to advance the steerable guide catheter (sgc) appears to be due to procedural circumstances; however, a definitive cause for the reported leak (loss of fluid column) could not be determined in this incident.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|