MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. EMERALD; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. EMERALD; SURGICAL ADJUNCTS Back to Search Results
Model Number 1CART30
Device Problem Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 06/25/2019
Event Type  malfunction  
Manufacturer Narrative
If implanted, if explanted, give date: not applicable, the cartridge is not an implantable device.Email address: unknown/not provided.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that there were 3 cartridges that appeared to have a sharp/warped, and melted tips, upon attempted implantation, the leading haptic broke/sheared off by the cartridge.There was no patient contact.No additional information provided.This report is for #1 of the 3 reported events.A separate report is being filed for each of the 3 cartridges.
 
Manufacturer Narrative
Device evaluation: the device was not returned at the manufacturing site; therefore; product testing could not be performed, and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.The search revealed that only one similar complaint for this production order number was received.Product deficiency was not identified.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERALD
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key8825035
MDR Text Key152216302
Report Number2648035-2019-00763
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474534469
UDI-Public(01)05050474534469(17)200124(10)CE00903
Combination Product (y/n)Y
PMA/PMN Number
K961242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 01/01/2005,11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/24/2020
Device Model Number1CART30
Device Catalogue Number1CART30
Device Lot NumberCE00903
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 06/25/2019
Initial Date FDA Received07/25/2019
Supplement Dates Manufacturer Received08/19/2019
10/25/2020
Supplement Dates FDA Received09/12/2019
11/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ZCB00 LENS SN: UNKNOWN; ZCB00 LENS SN: UNKNOWN
-
-