The returned product consisted of a solent dista thrombectomy catheter.The pump, effluent/supply line, shaft, tip, piston, and spike line were visually inspected.Blood was present outside and inside the device and 50ml in the attached waste bag, when received.Visual inspection of the device revealed that the shaft was twisted/kinked 30cm from the strain relief and that the shaft was detached 26cm proximal of the tip at the proximal and distal shaft transition.Microscopic examination revealed that there was a hole at the twisted/kinked location and the distal end that was detached was stretched with jagged edges.The damage to the both damaged areas indicates that there was force applied to the device during use.Functional testing was performed placing a test guidewire into the tip and advancing the device; however, there was resistance at the separated ends and it won't pass the twisted/kinked shaft after it passed the separated shaft.Inspection of the remainder of the device presented no other damage or irregularities.
|
Reportable based off on analysis completed on 7 july 2019.It was reported that the wire perforated the catheter wall outside the patient.An angiojet solent dista catheter was selected for a procedure in the patient's leg.The staff primed the catheter as usual, but when they inserted the wire into the catheter, they found that the wire came out of the side of the catheter.It appeared that the wire perforated the catheter wall and came out on the side of the shaft as if it were a rapid exchange device.This issue occurred outside of the patient.Another of the same device was used to complete the procedure.No patient complications were reported.The patient condition after the procedure was fine.However, device analysis revealed a detached shaft.
|