The stent and hemostasis valve tools were returned in the original pouch and vial.The covering is completely detached.There is blood on the covering.Four adhesive spots are visible on the covering.The stent has been crimped down, and the od of the stent is 0.133" the hemostasis valve tools are undamaged.The device history records were reviewed, and no issues were found.One device from every lot is tested for covering strength as part of the final inspection process.For lot number ccp-1046, the covering strength was 3.19008 which is above the acceptance criteria of: greater than 1.5 lbs.The root cause of the detachment cannot be determined.
|
Description of event as sent to numed by the distributor on the incident report - after stent was fully crimped and visually inspected, the physician inserted mounted covered cp stent into the appropriate tool.During this procedure, the physician observed that the ptfe on the stent was peel off while it was inserted the long sheath.The physician pulled out the covered mounted stent from the patient and used another stent.Another cvrd stent was implanted successfully.The patient discharged from the hospital.
|