RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
|
Back to Search Results |
|
Device Problem
Protective Measures Problem (3015)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2018.Date of report: 25jul2019.The manufacturer's field service engineer (fse) could not confirm the reported issue.The fse was able to identify during troubleshooting of the reported issue that the low-pressure alarm limits were not set correctly.This mdr has been reassessed as reportable after a request from the fda on march 1, 2019 to review complaints from 29 nov 2017 to 29 nov 2018.As this has been reassessed, it will appear to be a late mdr.
|
|
Event Description
|
The customer reported a low pressure regulation error.The unit was in clinical use at the time the reported issue was discovered; however, there was no harm to the patient or the user.
|
|