Catalog Number 4701000000
Device Problem Positioning Problem (3009)
Patient Problem No Patient Involvement (2645)
Event Date 04/01/2019
Event Type  malfunction  
Manufacturer Narrative
This mdr is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.The devices were evaluated in the field and the issue was confirmed; 2 units had broken/damaged components, 2 had rusted components, and 1 had a stripped component.The devices were repaired and returned.There was no remedial action taken.This device is not labeled for single use.
 
Event Description
This report summarizes 5 malfunction events, where it was reported the foot section or abduction assembly would not latch/lock.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LD304 MATERNITY BED
Type of Device
TABLE, OBSTETRICAL, AC-POWERED (AND ACCESSORIES)
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
mary klaver
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key8835998
MDR Text Key152534802
Report Number0001831750-2019-00775
Device Sequence Number1
Product Code HDD
UDI-Device Identifier07613327278095
UDI-Public07613327278095
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K950889
Number of Events Reported5
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number4701000000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/01/2019
Initial Date FDA Received07/29/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-