The patient was undergoing a thrombectomy procedure in the anterior tibial artery using a indigo system catrx aspiration catheter (catrx) from an indigo system catrx kit.During the procedure, the physician completed the first pass using the catrx.While attempting to perform the second pass, air was observed coming through the aspiration tubing (tubing).The physician withdrew the catrx from the 6f sheath and noted that the catrx was fractured proximal to the wire port.A syringe was connected to the luer lock and clot was flushed from both proximal end and distal tip of catrx.The procedure was completed using a new catrx, the same sheath and an indigo system cat6 aspiration catheter (cat6).There was no report of an adverse effect to the patient.
|
Results: the indigo system catrx aspiration catheter (catrx) was kinked approximately 114.0 cm from the hub.The guide wire lumen was not torn along the catheter shaft.Conclusions: evaluation of the returned catrx revealed a kinked device.If the device is forcefully mishandled during use, damage such as a kink may occur.During functional testing, the catrx was connected to a pump max and water was aspirated through the catrx.Bubbles were observed coming through the aspiration tubing during aspiration.The kink in the catrx lumen likely contributed to the reported air coming through the aspiration tubing during the procedure.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|