It was reported that the wire coating peeled off.A 330cm rotawire and 1.75mm rotalink plus were selected for use.During preparation, it was noted that the wire coating was peeled off when inserted into the burr outside patient's body.Subsequently, a large amount of coating material adhered on the tip of the rotaburr.The rotawire was replaced but the rotawires felt rough and the coating was uneven.Ablation was then stopped and the procedure was completed with the usual percutaneous coronary intervention (pci).No patient complications were reported.
|
Device evaluated by the manufacturer: the device wasreturned for analysis.The surface of the device was carefully inspected and some residues were found on some sections of it.Previous to performing the decontamination process, the device was inspected looking for the coating the issues reported by the customer.The surface looked in good condition, however, white residues were encountered on its surface and they looked like lumps.The surface of the device was inspected again and some residues were observed on the surface; however, the guidewire surface did not show abnormalities, it was in good condition.Besides, the device had the distal tip stretched.No more damages were found in the device.The dimensional inspection revealed that the overall length and the outer diameter (od) of the distal tip, middle of the device, and the proximal section were within specification.The directions for use indicate the following:"exercise care in handling of the guidewire during a procedure to reduce the possibility of accidental breakage, bending, kinking, or coil separation." "the guidewires are coated with a thin film of lubricant which may appear as a white powder.Do not wipe off lubricious coating.The lubricant can cause the guidewire to adhere to the inside of the tube, making unloading difficult.If this happens, gently tap the outside of the coil to loosen the wire.Use care not to stretch or damage the spring tip.".
|
It was reported that the wire coating peeled off.A 330cm rotawire and 1.75mm rotalink plus were selected for use.During preparation, it was noted that the wire coating was peeled off when inserted into the burr outside patient's body.Subsequently, a large amount of coating material adhered on the tip of the rotaburr.The rotawire was replaced but the rotawires felt rough and the coating was uneven.Ablation was then stopped and the procedure was completed with the usual percutaneous coronary intervention (pci).No patient complications were reported.
|