• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; LLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; LLD Back to Search Results
Model Number 518-067
Device Problem Material Separation (1562)
Patient Problems Low Blood Pressure/ Hypotension (1914); Iatrogenic Source (2498)
Event Date 07/16/2019
Event Type  Injury  
Manufacturer Narrative
Patient's weight was unavailable.Device lot number and expiration date are not available from the facility.The device was discarded by the user.Device manufacture date is dependent on the device lot number, thus is unavailable.The device was discarded.Therefore, no device evaluation can be performed.
 
Event Description
A philips representative reported that a cardiac lead management procedure commenced to remove a non-functional/fractured right ventricular (ev) implantable cardioverter defibrillator (icd) lead.A spectranetics lead locking device (lld) was utilized as the traction platform for the extraction.The lld snapped at the rv lead fracture area, which was at the proximal end of the coil.The patient's blood pressure dropped at that time.Rescue efforts, including cpr and sternotomy, revealed that the superior vena cava (svc) was torn during the extraction (please see mdr 1721279-2019-00136, as the svc tear is not attributed to the lld device).Rescue efforts commenced including sternotomy, and the lead was removed during open heart.Note: 4076 right atrial (ra) lead, rep believes, was taken out during the open heart along with the 0292 right ventricular (rv) lead, due to being potentially dislodged.Original intention was to just take out 0292 rv lead.The patient survived the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
catherine eaton
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key8852076
MDR Text Key152985204
Report Number1721279-2019-00137
Device Sequence Number1
Product Code DRB
UDI-Device Identifier00813132023065
UDI-Public00813132023065
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K140775
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,07/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number518-067
Device Catalogue Number518-067
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 07/16/2019
Initial Date FDA Received08/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BOSTON SCIENTIFIC 0292 ICD LEAD; MEDTRONIC 4076 PACING LEAD; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; SPECTRANETICS GLIDELIGHT LASER SHEATH; SPECTRANETICS TIGHRAIL ROTATING DILATOR SHEATH; SPECTRANETICS VISISHEATH DILATOR SHEATH
Patient Outcome(s) Required Intervention;
Patient Age23 YR
-
-