Model Number X SERIES |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
Death (1802)
|
Event Date 07/20/2019 |
Event Type
Death
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient in cardiac arrest (age & gender unknown),the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty mfc-to-cprd connector.The device was recertified and returned to the customer.The customer received a replacement mfc-to-cprd connector.Analysis for reports of this type has not identified an increase in trend.
|
|