Model Number X SERIES
Device Problem Unable to Obtain Readings (1516)
Patient Problem Death (1802)
Event Date 07/20/2019
Event Type  Death  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to treat a patient in cardiac arrest (age & gender unknown),the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
 
Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty mfc-to-cprd connector.The device was recertified and returned to the customer.The customer received a replacement mfc-to-cprd connector.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key8855818
MDR Text Key153115680
Report Number1220908-2019-02254
Device Sequence Number1
Product Code DRT
UDI-Device Identifier00847946006372
UDI-Public00847946006372
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K112432/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/24/2019
Initial Date FDA Received08/02/2019
Supplement Dates Manufacturer Received07/24/2019
Supplement Dates FDA Received09/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
-
-