MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number 9732719
Device Problems Problem with Software Installation (3013); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Patient Involvement (2645)
Event Date 07/10/2019
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: bi71000110r, serial/lot #: (b)(4).A medtronic representative went to the site to test and service the equipment.The mobile view station (mvs) computer displayed that 'bootmgr' was missing.The mvs computer was replaced.The system then passed the system checkout and was found to be fully functional.The mvs computer was returned to medtronic but was under analysis at the time of filing.Device manufacturing date unavailable.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging device.It was reported that a medtronic representative (rep) was working on the system and was unable to update the software.Additionally, the mobile view station (mvs) computer was no longer booting up.It was noted that ¿bootmgr¿ was missing.There was no patient involved.
 
Manufacturer Narrative
Device evaluation: analysis was completed for the mobile view station (mvs) computer.Functional testing, performance testing, and visual/physical examination were performed, and the reported problem was confirmed.The missing boot disk was confirmed.After a software install, however, the system was back to normal.It was determined that there was a software failure with the mvs computer with a failure mechanism of a corrupt operating system.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8857370
MDR Text Key153225925
Report Number3004785967-2019-01364
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9732719
Device Catalogue Number9732719
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2019
Initial Date Manufacturer Received 07/10/2019
Initial Date FDA Received08/02/2019
Supplement Dates Manufacturer Received09/05/2019
Supplement Dates FDA Received09/30/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-