MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
|
Back to Search Results |
|
Model Number 9732719 |
Device Problems
Problem with Software Installation (3013); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/10/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: other relevant device(s) are: product id: bi71000110r, serial/lot #: (b)(4).A medtronic representative went to the site to test and service the equipment.The mobile view station (mvs) computer displayed that 'bootmgr' was missing.The mvs computer was replaced.The system then passed the system checkout and was found to be fully functional.The mvs computer was returned to medtronic but was under analysis at the time of filing.Device manufacturing date unavailable.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging device.It was reported that a medtronic representative (rep) was working on the system and was unable to update the software.Additionally, the mobile view station (mvs) computer was no longer booting up.It was noted that ¿bootmgr¿ was missing.There was no patient involved.
|
|
Manufacturer Narrative
|
Device evaluation: analysis was completed for the mobile view station (mvs) computer.Functional testing, performance testing, and visual/physical examination were performed, and the reported problem was confirmed.The missing boot disk was confirmed.After a software install, however, the system was back to normal.It was determined that there was a software failure with the mvs computer with a failure mechanism of a corrupt operating system.If information is provided in the future, a supplemental report will be issued.
|
|