Model Number R SERIES |
Device Problem
Failure to Select Signal (1582)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device was unable to select energy level via external paddles.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was not returned to zoll medical corporation for evaluation; instead the paddles were returned.The returned paddles were subjected to extensive testing which included functional testing without duplicating the reported malfunction.The returned paddles were working as expected.Analysis for reports of this type has not identified an increase in trend.
|
|