• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Device Problems Electrical /Electronic Property Problem (1198); Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 05aug2019.
 
Event Description
The customer reported 3.3 v voltage rail failure.There was no patient or user harm reported.
 
Manufacturer Narrative
Date rec'd by mfr: 23oct2019.The device was evaluated by the field service engineer and the reported issue was confirmed.The field service engineer reset the connections to address the issue.The issue was resolved, the device was tested, and the device now functions as intended.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Event Description
The customer reported 3.3 v voltage rail failure.There was no patient or user harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8861875
MDR Text Key153432615
Report Number2031642-2019-05458
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/23/2019
Initial Date FDA Received08/05/2019
Supplement Dates Manufacturer Received07/23/2019
Supplement Dates FDA Received10/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-