• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Device Problem Computer Operating System Problem (2898)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 08/05/2019.
 
Event Description
The customer reported software corrupted.There was no patient involvement.
 
Manufacturer Narrative
Date received by manufacturer: 01nov2019.Date of report: 04nov2019.Device code: power problem.The manufacturer¿s international service technician confirmed the reported issue.3.3 v voltage rail failed error was observed.The manufacturer¿s international service technician reset all the internal connections of the device boards and cleaned the filter to address the reported problem.The unit successfully passed the required performance verification test.The determination could not be made that the device failed to meet specifications.The device was not being used for treatment when the reported event occurred, and there is a relationship of the device to the reported problem.Part was not returned to failure investigation (fi).The root cause cannot be determined until the device is returned and investigated.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8862041
MDR Text Key161952289
Report Number2031642-2019-05464
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/23/2019
Initial Date FDA Received08/05/2019
Supplement Dates Manufacturer Received07/23/2019
Supplement Dates FDA Received11/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-