• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. SPINAL ROD CUTTER MAXIMUM PIN DIAMETER; TRAUMA, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. SPINAL ROD CUTTER MAXIMUM PIN DIAMETER; TRAUMA, INSTRUMENT Back to Search Results
Catalog Number 00392500200
Device Problem Fracture (1260)
Patient Problem No Information (3190)
Event Date 06/26/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4) report source - (b)(6).It is unknown if product will be returning to zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the instrument fractured during the procedure.There was no reported patient or user injury.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual examination of the provided pictures noted the fractured jaw of the instrument, confirming the complaint.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual inspection of the returned cutter noted the following - there was an unidentified lot number etched on the pin cutter.One jaw is fractured.Visual inspection of both jaws shows that they exhibit heavy wear marks.There was a design feature missing on the returned product of the part number provided.Accordingly, zimmer part no.Was ruled out as the possible identity of the subject instrument.As product identification is not possible no further evaluations could be performed.The complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPINAL ROD CUTTER MAXIMUM PIN DIAMETER
Type of Device
TRAUMA, INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8863720
MDR Text Key154905670
Report Number0001822565-2019-03340
Device Sequence Number1
Product Code HXZ
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00392500200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/11/2019
Initial Date FDA Received08/06/2019
Supplement Dates Manufacturer Received10/29/2019
02/25/2021
Supplement Dates FDA Received10/30/2019
02/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-