MAUDE Adverse Event Report: SMITH & NEPHEW, INC. VIDEOARTHROSCOPE HD 4MM X 30 DEG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. VIDEOARTHROSCOPE HD 4MM X 30 DEG Back to Search Results
Catalog Number 72202961S
Device Problem Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/09/2019
Event Type  malfunction  
Event Description
It was reported that during surgery, the device had an halo in the lens.No backup device was available to complete the procedure, and no delay or patient injuries were reported.
 
Manufacturer Narrative
An evaluation was performed by the supplier and could confirm the customer complaint for there was a halo in the lens.A visual inspection was performed and showed the scope to have distal tip damage, a bent outertube and a broken sidearm.This damage is caused by contact with another source.No manufacturing related defects were observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOARTHROSCOPE HD 4MM X 30 DEG
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key8867938
MDR Text Key154570215
Report Number3003604053-2019-00086
Device Sequence Number1
Product Code HRX
UDI-Device Identifier00885554024203
UDI-Public00885554024203
Combination Product (y/n)N
PMA/PMN Number
K043395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 08/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72202961S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2019
Initial Date Manufacturer Received 07/10/2019
Initial Date FDA Received08/07/2019
Supplement Dates Manufacturer Received08/14/2019
Supplement Dates FDA Received08/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-