MAUDE Adverse Event Report: BATTLE CREEK EQUIPMENT THERMOPHORE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BATTLE CREEK EQUIPMENT THERMOPHORE Back to Search Results
Model Number 155
Device Problems Thermal Decomposition of Device (1071); Fire (1245); Sparking (2595); Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Customer stated, "pad sparked anf shot out flames." the product was returned for investigation.The customer did not claim any injury.An investigation was done to look into the customers complaint.The investigator found that the customer was twisting/wrapping the cord during use.This led to weakening of the cord and caused the burn the investigator found.This burn was likely the source of the sparks and flame the customer described.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE
Type of Device
THERMOPHORE
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
702 s reed rd
fremont IN 46737
Manufacturer Contact
uday bahal
MDR Report Key8868273
MDR Text Key153948382
Report Number1832415-2019-10624
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 07/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number155
Device Lot Number00155063193
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/07/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-