• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-15
Device Problem Partial Blockage (1065)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/10/2019
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).The affected part was requested to be returned to livanova (b)(4) for a detailed investigation.Corrective actions are in progress for this issue.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that the shaft angle encoder of a centrifugal pump system with tubing clamp intermittently locked and did not turn during procedure.There was no report of patient injury.
 
Manufacturer Narrative
Additional narrative: deviation could be reproduced during the investigation.The shaft encoder was stuck.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8872666
MDR Text Key175214037
Report Number9611109-2019-00607
Device Sequence Number1
Product Code DWA
UDI-Device Identifier04033817900986
UDI-Public(01)04033817900986(11)181009
Combination Product (y/n)N
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/10/2019
Initial Date FDA Received08/08/2019
Supplement Dates Manufacturer Received08/21/2019
Supplement Dates FDA Received09/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-