• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CAMERA HEAD LENS INT SYS; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CAMERA HEAD LENS INT SYS; ARTHROSCOPE Back to Search Results
Catalog Number 72203967
Device Problem No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/1901
Event Type  malfunction  
Event Description
It was reported that during shoulder scope, the device was whiting out.A backup device was available to complete the procedure with no delay or patient injuries.
 
Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of live video output malfunction was confirmed.Cause of video malfunction is a defective cable assembly.Camera head passed functional testing with a known good cable assembly installed.The complaint investigation has concluded that the root cause of malfunction is a defective cable assembly.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAMERA HEAD LENS INT SYS
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key8884341
MDR Text Key154000057
Report Number1643264-2019-00532
Device Sequence Number1
Product Code HRX
UDI-Device Identifier00885556629833
UDI-Public00885556629833
Combination Product (y/n)N
PMA/PMN Number
K153606
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 09/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72203967
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2019
Initial Date Manufacturer Received 08/07/2019
Initial Date FDA Received08/12/2019
Supplement Dates Manufacturer Received09/24/2019
Supplement Dates FDA Received09/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-