• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC UNK - VEPTR IMPLANTS; PROSTHESIS, RIB REPLACEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC UNK - VEPTR IMPLANTS; PROSTHESIS, RIB REPLACEMENT Back to Search Results
Device Problem Migration (4003)
Patient Problems Pain (1994); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
Brand name, common device name, procode, mfr, lot #, part #, udi #, 510k: this report is for an unknown veptr implants/unknown lot.Part and lot number are unknown; udi number is unknown.Device available for evaluation: complainant part is not expected to be returned for manufacturer review/investigation.Device evaluated by mfr, manufacture date: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
This report is being filed after the review of the following journal article: parnell, s.E.Et al (2015), vertical expandable prosthetic titanium rib (veptr): a review of indications, normal radiographic appearance and complications, pediatric radiology, vol.45 (4), pages 606-616 (usa).The aim of this study is to reviews the clinical indications and strategies for the use of the veptr device, imaging modalities that may be used in the evaluation and management of patients with thoracic insufficiency syndrome and familiarizes the reader with the types of complications that can occur with emphasis on the radiographic findings that are seen with these complications.An unknown number of patients were included in the study.Surgery was performed using vertical expandable prosthetic titanium rib (veptr; synthes, west chester, pa).The following complications were reported: an (b)(6)-year-old boy had right posterior rib fracture adjacent to the superior cradle of the veptr device.A (b)(6)-year-old girl had migration of the superior rib cradle and erosion through the ribs.An (b)(6)-year-old girl had inferolateral dislodgement of the s-hook and increased thoracolumbar levoscoliosis.An (b)(6)-year-old girl had erosion of hardware (pelvic hook) through the skin with persistent serosanguinous drainage from the wound, raising question of infection.An (b)(6)-year-old girl had abnormal migration of the lumbar hook into the spinal canal at the lumbosacral junction causing soft-tissue damage and radiculopathy.A (b)(6)-year-old boy had migration of the superior cradle over time indicating dislodgement of the cradle from the original rib anchors.A (b)(6)-year-old girl had a fracture (broken hardware) of the inferior portion of the rib sleeve of a hybrid veptr device.This report is for an unknown synthes veptr implants.This is report 5 of 7 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK - VEPTR IMPLANTS
Type of Device
PROSTHESIS, RIB REPLACEMENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key8889460
MDR Text Key154260488
Report Number2939274-2019-59809
Device Sequence Number1
Product Code MDI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type literature
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/17/2019
Initial Date FDA Received08/13/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age8 YR
-
-