• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 2 8MM; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 2 8MM; TEMPLATE Back to Search Results
Catalog Number 00584102208
Device Problem Insufficient Information (3190)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/26/2019
Event Type  malfunction  
Event Description
It was reported that during surgery the zuk insert trial broke.Nothing got in patient.No harm to patient or staff.No letter required at this time.No delay was reported and a backup device was available.The device broke over the patient's incision.
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device fractured in half and both pieces were returned.The device also had numerous nicks and gouges in the surface.The device was manufactured in 2014 and exhibits signs of extensive wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE PROVISIONAL SZ 2 8MM
Type of Device
TEMPLATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8892465
MDR Text Key154261743
Report Number1020279-2019-03050
Device Sequence Number1
Product Code HWT
UDI-Device Identifier00885556619384
UDI-Public00885556619384
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00584102208
Device Lot Number6279217
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2019
Initial Date Manufacturer Received 07/26/2019
Initial Date FDA Received08/14/2019
Supplement Dates Manufacturer Received07/26/2019
Supplement Dates FDA Received09/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-