LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/26/2019
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that on a s5 double head pump an error message was displayed and there was no rotation.After a restart, the pump worked as expected.This event occured during procedure.There was no report of patient injury.
 
Manufacturer Narrative
H.10: a complaints database research reveled that the reported event was a re-occurrence of the already reported event (initial medwatch # mw006216, follow up medwatch # mw006670).Based on the follow up communication and the result of the investigation of the previous complaint, it is likely that the cause can be related to electromagnetic interferences in the operation theatre.As per the instruction for use, it is recommended to make sure that no electromagnetic interference is caused by third party devices.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Hold dpm 6.13.23.See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8892641
MDR Text Key175858805
Report Number9611109-2019-00633
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/27/2019
Initial Date FDA Received08/14/2019
Supplement Dates Manufacturer Received12/23/2019
Supplement Dates FDA Received01/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-