DENTSPLY LLC CAVITRON JET SPS ULTRASONIC SCALER AND AIR POLISH SYSTEM; SCALER, ULTRASONIC Back to Search Results
Model Number G120
Device Problems Restricted Flow rate (1248); Overheating of Device (1437); Improper Flow or Infusion (2954)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a g120 scaler, there was poor water flow and the unit was running hot; no injury resulted.
 
Manufacturer Narrative
Handpiece cable water line too short and getting pinched off causing no water flow.Air system flooded; debris in water system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON JET SPS ULTRASONIC SCALER AND AIR POLISH SYSTEM
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
MDR Report Key8895001
MDR Text Key158896747
Report Number2424472-2019-00127
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
K970342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG120
Device Catalogue NumberG120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2019
Initial Date Manufacturer Received 07/15/2019
Initial Date FDA Received08/14/2019
Supplement Dates Manufacturer Received07/15/2019
Supplement Dates FDA Received09/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-