• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number 8900
Device Problems Failure to Advance (2524); Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/18/2019
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older device evaluated by mfr: the tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed 2 kinks located at the occ handle and 177.5cm from the tip.There was coating peeling from the 177.5 location.The occ handle was connected to the ffr link for signal verification.The signal was present as designed.The occ handle cap was loosened to remove the wire.There was no issue with removing the wire.The sensor port showed no residue of body fluids.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The complaint of a kink was confirmed.The pressure wire was connected to the analysis support test bench and all applicable data was correct as designed, there was no difficulty in connecting the wire to the test equipment.The coefficient was confirmed to be in specification.
 
Event Description
It was reported that difficulty crossing the lesion and a torque problem occurred.The 90% stenosed target lesion was located in the tortuous distal right coronary artery (rca).A comet pressure guidewire was used but was unable to cross the tortuous area in the mid rca.It was noticed that the torque could not be managed at all.The procedure was completed with a non bsc device that was able to cross the lesion.No patient complications were reported in relation to this event.However, returned device analysis revealed peeled coating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key8902226
MDR Text Key154582245
Report Number2134265-2019-09313
Device Sequence Number1
Product Code DXO
UDI-Device Identifier08714729904403
UDI-Public08714729904403
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 08/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/10/2021
Device Model Number8900
Device Catalogue Number8900
Device Lot Number0023329820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/29/2019
Initial Date FDA Received08/16/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-