BOSTON SCIENTIFIC CORPORATION FLEXTOME CUTTING BALLOON; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
|
Back to Search Results |
|
Model Number 3822 |
Device Problem
Break (1069)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/28/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
It was reported that the delivery shaft fractured.A 10/2.75 flextome cutting balloon was selected for use.During preparation, when the device was being unpacked, it was noted that the cutting balloon's delivery shaft was fractured, and the device could not be used.The procedure was completed with another of the same device.No complications reported, and the patient was stable.
|
|
Event Description
|
It was reported that the delivery shaft fractured.A 10/2.75 flextome cutting balloon was selected for use.During preparation, when the device was being unpacked, it was noted that the cutting balloon's delivery shaft was fractured, and the device could not be used.The procedure was completed with another of the same device.No complications reported, and the patient was stable.
|
|
Manufacturer Narrative
|
Initial reporter phone: (b)(6).Device evaluated by the manufacturer: the device was returned for analysis.A visual and microscopic examination was performed on the returned flextome cb monorail 10/2.75.The balloon wings were relaxed.A visual and microscopic examination found no issues with the balloon or blades.The tip section of the device was visually and microscopically examined, and no issues were noted with its profile.A visual and microscopic examination found no issue with the profile of the device markerbands.A visual and tactile examination identified a hypotube break at approximately 85 cm from the distal end of strain relief.Multiple hypotube kink were also identified along several locations of the hypotube shaft.The type of damages noted are consistent with excessive force that could have been applied to the delivery system.No issues were noted on the shaft polymer extrusion shaft.No other issues were identified during the product analysis.The directions for use states under preparation section that care should be taken not to kink or bend the shaft.
|
|
Search Alerts/Recalls
|
|
|