• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALEM MEDICAL LTD. MALEM ENURESIS ALARM; ALARM, CONDITIONED RESPONSE ENURESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALEM MEDICAL LTD. MALEM ENURESIS ALARM; ALARM, CONDITIONED RESPONSE ENURESIS Back to Search Results
Model Number PRO
Device Problem Overheating of Device (1437)
Patient Problems Burn(s) (1757); Pain (1994)
Event Date 07/25/2019
Event Type  Injury  
Event Description
The dr in our clinic prescribed an electronic enuresis alarm for pt who suffers from chronic incontinence.Pt is immobile and needs medical attention.The pt's purchased malem enuresis alarm (pro) model to detect urine leaks.Child was wearing alarm and awake when incident occurred.It was the first use case of the device.Within less than an hour, system developed a flaw which caused the temperature to rise significantly, causing child's skin to burn.She was awake and in pain from burns.Pt was brought back to clinic.The injuries are the result of extreme heat and hot plastic touching pt's skin.Burns have been treated and pt has returned home.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALEM ENURESIS ALARM
Type of Device
ALARM, CONDITIONED RESPONSE ENURESIS
Manufacturer (Section D)
MALEM MEDICAL LTD.
MDR Report Key8915497
MDR Text Key155195602
Report NumberMW5089207
Device Sequence Number1
Product Code KPN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date 08/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberPRO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/20/2019
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age4 YR
-
-