Model Number 8886848813
Device Problem Device Slipped (1584)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 07/30/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during laparoscopic radical resection of rectal cancer, clip detachment from the tissue was noted, which resulted from insufficient clamping force.Another clip applier (absorbable) was used to complete the surgery.
 
Event Description
According to the reporter, during laparoscopic radical resection of rectal cancer, clip detachment from the tissue was noted, which resulted from insufficient clamping force.Another clip applier (absorbable) was used to complete the surgery.There was no patient injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led three video evaluations of one device.A visual inspection of the returned videos noted: the first video depicts a laproclip applied to tissue and the customer using forceps to try and remove the clip.The second video depicts two laproclips applied to the card board container and then the customer pulls one off by hand and then attempt to remove the other by hand.The third video depicts a laproclip being applied to tissue and then pulled off with forceps.The videos were reviewed and no abnormalities were observed with the product.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPRO-CLIP
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8917793
MDR Text Key155114447
Report Number1219930-2019-04830
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier20884521057668
UDI-Public20884521057668
Combination Product (y/n)N
PMA/PMN Number
K925602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Model Number8886848813
Device Catalogue Number8886848813
Device Lot NumberN9A0255Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2019
Initial Date Manufacturer Received 07/30/2019
Initial Date FDA Received08/21/2019
Supplement Dates Manufacturer Received09/10/2019
10/07/2019
Supplement Dates FDA Received10/02/2019
10/23/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age89 YR
Patient Weight51
-
-