Catalog Number 227456000
Device Problems Device-Device Incompatibility (2919); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Not Applicable (3189)
Event Date 08/06/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the drill bit (p/n: 227456000) and the drill shaft (p/n: 227452000) were able to join during the tha surgery of the patient¿s left hip joint on (b)(6) 2019 however the lock did not function and could not be fixed even if the shaft was turned to the lock position.The surgery was completed within a 30 minutes surgical delay and there was no adverse consequence to the patient.No further information is available.
 
Manufacturer Narrative
This product was reported in error.No patient death, serious injury or evidence of reportable product malfunction occurred.Depuy considers the investigation closed at this time.Should any additional information be received to change the outcome of the performed investigation, the complaint will be re-opened.Product complaint #: (b)(4).Investigation summary: examination of the returned instrument found no defect with the instrument.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint#: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Corrected : h6.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET 3.8 DIMTR DR BIT 25MM
Type of Device
HIP INSTRUMENTS : DRILLS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key8930333
MDR Text Key155556112
Report Number1818910-2019-102163
Device Sequence Number1
Product Code HTW
UDI-Device Identifier10603295109310
UDI-Public10603295109310
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 08/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number227456000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2019
Initial Date Manufacturer Received 08/05/2019
Initial Date FDA Received08/26/2019
Supplement Dates Manufacturer Received09/04/2019
10/29/2019
10/31/2019
Supplement Dates FDA Received09/24/2019
10/30/2019
11/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-