• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY ACROBAT® 2 CALIBRATED TIP WIRE GUIDE; OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY ACROBAT® 2 CALIBRATED TIP WIRE GUIDE; OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number G47617
Device Problems Peeled/Delaminated (1454); Off-Label Use (1494)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2019
Event Type  malfunction  
Manufacturer Narrative
Mtw ercp catheter (0130311, standard metal tip, 0.035 inch/mtw (abis).Investigation evaluation: the wire guide tip is bent approximately 1.5 cm, 3.0 cm, and 5.0 cm from the distal end.Approximately 4.8 cm to 5.7 cm from the distal end is a section of bare core wire.A section of the coating approximately 0.9 cm long is frayed and hanging from the wire guide, the coating still attached at approximately 4.8 cm from the distal end.Rough surfaces are found throughout the entire length of the wire guide.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the report states that this incident occurred while using this product with an metal tip device.This is the most likely cause for the reported observation.The instructions for use for this product line caution the user: "use of this wire guide with metal tip ercp devices may result in damage to the external coating and/or tip of the wire guide".Prior to distribution, all acrobat® 2 calibrated tip wire guides are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.Additional comments regarding this report: based on the information provided that the device was used off label with a metal tip device, a cook representative has been requested to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook acrobat® 2 calibrated tip wire guide.The acrobat® 2 wire guide was combined with another manufacturer's metal tip catheter prior to insertion into endoscope.Then, the two devices as a unit were advanced through the endoscope to the papilla for ercp.The user was exploring the bile duct with manipulating the wire guide several times in order to insert the unite devices into the bile duct though, he failed.He removed only the wire guide to change the ercp catheter, then he noticed that the coating material of the wire guide was peeled-off at approximately 5 cm from the tip.Another acrobat® 2 wire guide was used instead.There have been no adverse effects to the patient reported.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT® 2 CALIBRATED TIP WIRE GUIDE
Type of Device
OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key8931399
MDR Text Key155635628
Report Number1037905-2019-00497
Device Sequence Number1
Product Code OCY
UDI-Device Identifier00827002476172
UDI-Public(01)00827002476172(17)220411(10)W4202526
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/11/2022
Device Model NumberG47617
Device Catalogue NumberAWG2-35-450-A
Device Lot NumberW4202526
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/01/2019
Initial Date FDA Received08/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-