• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955467
Device Problem Misassembled (1398)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/24/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided pictures showed that the pbp and effluent lines were incorrectly assembled on the support plate.The reported condition was confirmed.The cause is an error during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with a prismaflex m150 set, the replacement line was crossed with the citric acid line.There was patient involvement but no patient impact and no medical intervention associated with this event.No additional information is provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M150
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - MEYZIEU
meyzieu cedex rhone
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8931821
MDR Text Key193421421
Report Number8010182-2019-00174
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K080519
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/01/2020
Device Catalogue Number955467
Device Lot Number18J1203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/29/2019
Initial Date FDA Received08/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-