• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 29901
Device Problems Operating System Becomes Nonfunctional (2996); Unexpected Shutdown (4019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/31/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer shut down while testing.The programmer is expected to be returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis:analysis was unable to confirm the customer comment that the programmer shut down while testing.Device powers up and interrogates without shutting down.Found multiple errors in the error log.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the screen would go black while checking devices.The programmer would reboot and not finish the interrogation.The programmer was returned for service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK ENCORE PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8936595
MDR Text Key174940183
Report Number3004593495-2019-00878
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00643169369030
UDI-Public00643169369030
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number29901
Device Catalogue Number29901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/31/2019
Initial Date FDA Received08/27/2019
Supplement Dates Manufacturer Received08/23/2019
09/11/2019
Supplement Dates FDA Received09/06/2019
09/12/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-