Model Number 29901 |
Device Problems
Operating System Becomes Nonfunctional (2996); Unexpected Shutdown (4019)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/31/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the programmer shut down while testing.The programmer is expected to be returned for service.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product analysis:analysis was unable to confirm the customer comment that the programmer shut down while testing.Device powers up and interrogates without shutting down.Found multiple errors in the error log.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that the screen would go black while checking devices.The programmer would reboot and not finish the interrogation.The programmer was returned for service.
|
|