RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6).Date of report: 27aug2019.The customer is using products not tested and validated to be used on the v60.Bacteria filter: hudson rci bacterial/viral filter.Teleflex circuit: hudson rci adult conchasmart noninvasive breathing circuit with conchasmart column.Teleflex heater: hudson rci neptune philips clinical specialist informed cu the hudson heated circuit has too much resistance and is not approved for the v60.Only the f&p 25 mm circuit for heated wire is approved at this time.No parts were returned for failure investigation; therefore, the root cause at the component level could not be determined.
|
|
Event Description
|
The customer reported a few units that were not going in to standby.The device was in use at the time of the event; however, there was no patient harm.
|
|