Model Number 407201
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Date 08/22/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported cardiac perforation may have been procedure related.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Event Description
During a pulmonary vein isolation ablation procedure, a pericardial effusion occurred.The transseptal puncture was noted to be difficult and the guidewire was seen to advance into the pericardial space on fluoroscopy and echocardiogram.Contrast was injected and moderate staining in the superior pericardium was seen.The patient remained stable and no intervention was required but the procedure was cancelled.There were no performance issues with any abbott device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8950721
MDR Text Key156134298
Report Number3008452825-2019-00433
Device Sequence Number1
Product Code DRC
UDI-Device Identifier05414734205160
UDI-Public05414734205160
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K122587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model Number407201
Device Catalogue Number407201
Device Lot Number7015657
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/22/2019
Initial Date FDA Received08/30/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-