• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JABIL CIRCUIT(SHANGHAI) LTD BIS; ELECTRODE, CUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JABIL CIRCUIT(SHANGHAI) LTD BIS; ELECTRODE, CUTANEOUS Back to Search Results
Model Number 186-0106
Device Problem High Readings (2459)
Patient Problem No Patient Involvement (2645)
Event Date 08/08/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the unit's reading was above the higher limit 100.There was no patient involvement.
 
Manufacturer Narrative
Additional information: d10 (return date), g4, h3, h6 h3.Evaluation summary: nine devices were received and evaluated.All sensors passed inspection.An impedance check was performed.All electrodes passed the impedance check.The investigation found the devices to function normally.No new formal investigation is required.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIS
Type of Device
ELECTRODE, CUTANEOUS
Manufacturer (Section D)
JABIL CIRCUIT(SHANGHAI) LTD
no 600 tian lin road
shanghai,20 20023 3
CN  200233
MDR Report Key8953420
MDR Text Key156717827
Report Number2936999-2019-00626
Device Sequence Number1
Product Code GXY
UDI-Device Identifier20884521134307
UDI-Public20884521134307
Combination Product (y/n)N
PMA/PMN Number
K093183
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/08/2020
Device Model Number186-0106
Device Catalogue Number186-0106
Device Lot Number0508191L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2019
Initial Date Manufacturer Received 08/08/2019
Initial Date FDA Received09/02/2019
Supplement Dates Manufacturer Received09/20/2019
Supplement Dates FDA Received10/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-