Model Number 3824 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/30/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that the balloon detached from the catheter.A 15/4.00 flextome cutting balloon was selected for use.Upon removal of the plastic shield from the balloon's end, it was noted that the balloon separated from the catheter.The device could not be used.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.The balloon protector of the device was not returned for analysis.A visual examination identified that the balloon material was completely detached from the device at both distal and proximal balloon bonds.The balloon material, including all blades, was not returned for analysis.A microscopic examination identified that the shaft was kinked at more than one location.A visual and microscopic investigation identified no damage or any issues with the markerbands of the device that could have contributed to the complaint incident.A visual and tactile examination found the hypotube of the device to be kinked at more than one location.The type of damage noted is consistent with excessive force being applied to the device.No other issues were identified during the product analysis.
|
|
Event Description
|
It was reported that the balloon detached from the catheter.A 15/4.00 flextome cutting balloon was selected for use.Upon removal of the plastic shield from the balloon's end, it was noted that the balloon separated from the catheter.The device could not be used.There were no patient complications reported.
|
|
Search Alerts/Recalls
|