ZIMMER BIOMET, INC. TRABECULAR METAL BASE PLATE 15 MM POST LENGTH; PROSTHESIS, SHOULDER Back to Search Results
Catalog Number 00434901500
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bone Fracture(s) (1870); Pain (1994); Osteopenia/ Osteoporosis (2651); Limited Mobility Of The Implanted Joint (2671); Patient Problem/Medical Problem (2688)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: 00434901413, humeral stem 14 mm, 6248547; 00434903603, poly liner plus 3 mm, 62471628; 00434903611, glenosphere 36 mm diameter, 62540042.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2019-03681, 0001822565-2019-03682, 0001822565-2019-03683.
 
Event Description
It was reported through the clinical study that the patient had an initial right reverse total shoulder replacement.Subsequently, the patient complained of dissatisfaction and pain due to acromion fracture six (6) months post primary implantation after an unknown incident in therapy and the patient was treated to full recovery.No additional patient harm has been reported.
 
Event Description
No additional event information is available at the time of this report.
 
Manufacturer Narrative
His follow-up report is being submitted to relay additional information.Reported event was confirmed by review of medical records.Device history record (dhr) was reviewed and no discrepancies were found.Review of provided ct scan identified chronic ununited fracture of the acromion process with inferiorly displace fracture fragment narrowing the subacromial space.Fracture of acromion of scapula, right, sequela.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRABECULAR METAL BASE PLATE 15 MM POST LENGTH
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8956790
MDR Text Key156347948
Report Number0001822565-2019-03684
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
PMA/PMN Number
K121543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Type of Report Initial,Followup
Report Date 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number00434901500
Device Lot Number62541526
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/12/2019
Initial Date FDA Received09/03/2019
Supplement Dates Manufacturer Received09/13/2019
Supplement Dates FDA Received09/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age73 YR
Patient Weight78
-
-