Catalog Number 71751108
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/13/2019
Event Type  malfunction  
Event Description
It was reported that during surgery guide bolt is stripped.No harm to patient or staff.5 minute procedural delays.Smith & nephew back-up was not available.
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The threads on both ends of the device are damaged, rendering them inoperable.The hex end is also heavily damaged from impacts.The device was manufactured in 2004 and exhibits signs of extensive wear/usage.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT NAIL GUIDE BOLT
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8958231
MDR Text Key156391291
Report Number1020279-2019-03260
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71751108
Device Lot Number472419A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2019
Initial Date Manufacturer Received 08/13/2019
Initial Date FDA Received09/03/2019
Supplement Dates Manufacturer Received08/13/2019
Supplement Dates FDA Received09/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-