• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ADAPTA; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ADAPTA; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number ADDRL1
Device Problems Electrical /Electronic Property Problem (1198); Pacing Problem (1439); Pocket Stimulation (1463)
Patient Problems Muscle Stimulation (1412); Undesired Nerve Stimulation (1980); Pain (1994); Burning Sensation (2146)
Event Date 08/11/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient then experienced muscle stimulation, painful pocket stimulation and a burning sensation near the sight of the device after an accidental impact to the implantable pulse generator (ipg).The device was reported to have mode switched and exhibited a high-output.It was also noted that right ventricular (rv) lead exhibited high thresholds, intermittent non-capture and that a polarity switch had occurred.The device was explanted, the rv lead was capped and both were replaced.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.The battery indicator signifying that it is time for device replacement occurred after explant.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTA
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8959894
MDR Text Key156425294
Report Number3004209178-2019-16920
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier00613994222084
UDI-Public00613994222084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/28/2016
Device Model NumberADDRL1
Device Catalogue NumberADDRL1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/11/2019
Initial Date FDA Received09/04/2019
Supplement Dates Manufacturer Received09/13/2019
Supplement Dates FDA Received09/27/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
383059 LEAD, 5076-45 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age21 YR
-
-