Catalog Number A066100000300 |
Device Problem
Break (1069)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it was reported that a start-x tip broke during use.The event outcome is unknown as of this mdr evaluation.
|
|
Manufacturer Narrative
|
The involved start-x tip ems inserts 3 that broke during use were not returned and cannot be analyzed.Nothing unusual to report was found during dhr review (batch #1508351).We cannot rule on the compliance of the power settings selected by the customer in comparison with the ones recommended by maillefer which are exclusively given for ems generators.Root causes are not identified.We will track this kind of event and monitor the trend.Generally speaking, several other factors may contribute to breakage issue, such as usury, pressure or incorrect technique.All of these factors may affect the longevity of the tip.For information, we noted that the returned device had a worn active part.Multiple unsuccessful attempts were made to obtain the patient outcome.
|
|
Search Alerts/Recalls
|