• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPIC CO2 REGULATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPIC CO2 REGULATION UNIT Back to Search Results
Model Number UCR
Device Problem Pressure Problem (3012)
Patient Problems Pain (1994); Bowel Perforation (2668)
Event Type  Injury  
Event Description
The manufacturer became informed that 24 hours after a diagnostic colonoscopy procedure in (b)(6) 2019, a patient experienced pain and was inspected.The doctor at the user facility indicated that the patient was determined to have a perforated rectum and the doctor believed the injury to the patient was caused by over pressure from the endoscopic co2 regulation unit and gas/water valve.The patient was admitted to a different hospital/surgeon and required a exploratory laparotomy.The patient was hospitalized for a few weeks and discharged as the patient also had a pre-existing condition of severe arterial vascular disease in the lower extremities that eventually required amputation of one his legs.The technicians at the user facility performed a functional inspection prior to procedure and there were no anomalies noted.The doctor noticed that there was pressure being lost at the time but wasn't sure what it was, but the doctor later believed it to be pressure being lost from the perforation to the patient.This is for report 1 of 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPIC CO2 REGULATION UNIT
Type of Device
CO2 REGULATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key8961003
MDR Text Key157885697
Report Number2951238-2019-01100
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170239113
UDI-Public04953170239113
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Type of Report Initial
Report Date 09/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUCR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA09/04/2019
Distributor Facility Aware Date08/07/2019
Event Location Hospital
Date Report to Manufacturer08/07/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-