• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HAR9F
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2019
Event Type  malfunction  
Manufacturer Narrative
The facility stated the device was not saved.As the device was not returned for evaluation, inspection was unable to be performed.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore the most likely root cause(s) are: incidental contact with device and external instrument causing damage to the tissue pad too much force or torque applied to instrument, or grasping/pulling applying pressure between instrument blade and tissue pad without having tissue between them the instructions for use (ifu) state: care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument.If this happens, there may be a system failure signaled by a continuous tone or alert screen when either of the foot pedals or hand control buttons is depressed.Keep the jaws of the device open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad and increased blade, clamp arm, and distal shaft temperatures.Thermal influences such as fluids or minimal to no tissue in the jaws may affect the presence or timing of the tone change.Blood and tissue buildup between the blade and shaft may result in abnormally high temperatures at the distal end of the shaft.To prevent burn injury, remove any visible tissue buildup.At the distal end of the shaft.During and following activation in tissue, the instrument blade and clamp arm may become hot.Avoid unintended contact with tissue, drapes, surgical gowns, or other unintended sites at all times.Incidental and prolonged activation against solid surfaces, such as bone, may result in blade heating and subsequent blade failure, and should be avoided.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
 
Event Description
It was reported that the white tip fell off of one side of the harmonic scalpel (har9f) while in use.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Manufacturer Narrative
Additional information received from the customer after the initial mdr was filed was added to the executive summary.This information does not change the type of reportable event.
 
Event Description
It was reported that the white tip fell off of one side of the harmonic scalpel (har9f) while in use.The tip did not fall into the patient.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
MDR Report Key8961790
MDR Text Key158966076
Report Number0001056128-2019-00043
Device Sequence Number1
Product Code NLQ
UDI-Device Identifier07613327348248
UDI-Public07613327348248
Combination Product (y/n)N
PMA/PMN Number
K170456
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/02/2022
Device Model NumberHAR9F
Device Catalogue NumberHAR9FRR
Device Lot Number10269208
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/08/2019
Initial Date FDA Received09/04/2019
Supplement Dates Manufacturer Received09/06/2019
Supplement Dates FDA Received09/09/2019
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
-
-