DEPUY ORTHOPAEDICS INC US ALTRX NEUT 36IDX58OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 122136058
Device Problem Difficult to Insert (1316)
Patient Problem Not Applicable (3189)
Event Date 08/16/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during a routine left total hip replacement, a 58mm pinnacle socket was inserted.During impaction of the 58mmx36mm n altrex liner, it was noted that the liner would not seat correctly.The liner was removed, and subsequently a new liner was inserted.It is noted the second liner was seated appropriately and impacted.Surgery prolonged of 15 minutes.Doe: (b)(6) 2019; left hip.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: examination of the returned liner is unable to identify any product error or root cause for the problem experienced.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX NEUT 36IDX58OD
Type of Device
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8964650
MDR Text Key156585666
Report Number1818910-2019-103163
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295016359
UDI-Public10603295016359
Combination Product (y/n)N
PMA/PMN Number
K062148
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number122136058
Device Lot NumberJ4093Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2019
Initial Date Manufacturer Received 08/16/2019
Initial Date FDA Received09/05/2019
Supplement Dates Manufacturer Received10/11/2019
Supplement Dates FDA Received10/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age70 YR
-
-