Model Number MDT-ICD
Device Problem Electromagnetic Interference (1194)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/29/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was hearing an alert.The device was interrogated and showed that no condition met tone had occurred.The patient is noted to be implanted with a left ventricular assist device (lvad).It is suspected that the alerts were due to the magnets used to keep the lvad system in place.The device remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Corrected: this event is a duplicate of fda report number 9614453-2019-02742.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC ICD
Type of Device
PULSE-GENERATOR, PACEMAKER, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8965981
MDR Text Key156631496
Report Number2182208-2019-01609
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMDT-ICD
Device Catalogue NumberMDT-ICD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/29/2019
Initial Date FDA Received09/05/2019
Supplement Dates Manufacturer Received09/10/2019
Supplement Dates FDA Received09/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-