The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pto leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.Kit lot h318 was reviewed.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot h318 shows no trends.Trends were reviewed for complaint category, pto leak.No trends were detected for this complaint category.The complaint kit, smart card, and a photograph were returned for evaluation.The customer supplied photograph shows a blood leak on the pump deck near the return pump.A review of the returned smart card data verified that the treatment was successfully completed and blood was returned to the patient.Examination of the returned kit found dried blood on the lower right hand corner of the pto.The kit was pressure tested to check for leaks, and no leaks were found.The reported blood leak is verified based on the customer provided photograph and dried blood observed within the pto.However, the exact source of the leak could not be identified as pressure testing of the returned kit did not identify any leaks.As a result, a root cause for the pto leak could not be determined based on the available information.No further action required at this time.This investigation is now complete.Mc: (b)(4).B.K.: (b)(6) 2019.
|
The customer called to report that they experienced a pump tubing organizer (pto) leak with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer stated the blood leak was observed near the return pump after approximately 1500 ml of whole blood had been processed.In addition, the customer stated there were no cellex instrument alarms prior to the pto leak.The patient was reported to be in stable condition.A photograph of the complaint issue was provided by the customer for evaluation.In addition, the complaint kit, with smart card, was returned for evaluation.
|