Model Number 16402
Device Problem Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/13/2019
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) verified that the pump was displaying "failure 5." he replaced the roller pump central processing unit (cpu)/control printed circuit board assembly (pcba).The device operated to the manufacturer's specifications.The suspect device was sent back to the manufacturer for further evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the pump was displaying "failure 5".As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Updated block: h6.The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed that the central processing unit (cpu) control board was reporting a 'failure 5' error message during operation.He observed the cpu control board to fail six out of 10 times.When replacing the cpu with a lab use only board the pump function as intended, indicating that the user facility's cpu control board was bad.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 UNIVERSAL ROLLER PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key8971078
MDR Text Key156812232
Report Number1828100-2019-00476
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 11/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/13/2019
Initial Date FDA Received09/06/2019
Supplement Dates Manufacturer Received09/06/2019
11/01/2019
Supplement Dates FDA Received09/26/2019
11/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-