A steris service technician arrived onsite following the reported event to inspect the 5085 surgical table and found the table to be operating according to specification.No repairs were required.The table was returned to service.The technician was informed that as the employee was re-positioning the patient, the pad on the surgical table shifted causing the employee to "sprain" their lower back.The 5085 surgical table operator manual states (pg 4-3): "place pads onto tabletop by aligning with sides of table and pressing into place on fastening strips." the 5085 surgical table operator manual also states (pg 5-28): "place pads in position and firmly press hook-and-loop strips together.While onsite, the technician counseled user facility personnel on the proper use and operation of the 5085 surgical table.No additional issues have been reported.
|